The ozone analyzer is developed for ambient air quality monitoring. Based on UV absorption technology and advanced microprocessor technology, it provides precise and reliable measurements of O3 in the range of μmol/mol-umol/mol (ppb to ppm).
This product offers high accuracy, reliability, quick response, easy operation, and broad applicability. The analyzer supports both manual and automatic calibration, with customizable settings. It also allows for remote control and data upload. Real-time measurements and instrument status can be automatically transmitted and accessed, enabling user to monitor O3 instantly and accurately.
· UV absorption method measurement, meeting EPA standards.
· Measurement range: 0-500 ppb and 0-20ppm, adjustable by the user.
· Microprocessor-driven for multi-functional operation.
· Continuous self-check with alarm function.
· Light source intensity self-check.
· Bi-directional RS232 port for remote control.
· Digital output for operating parameters.
· Adaptive signal filtering for faster response.
· Temperature and pressure compensation.
· Color touchscreen display.
· Large memory for automatic historical data storage.
Parameters | Specification |
Range | Min: 0-500 nmol/mol (ppb) Max: 0-20 umol/mol (ppm) |
Units | nmol/mol (ppb),mol/mol (ppm),g/m3, mg/m2 ( Optional) |
Zero Noise | ≤ 1 nmol/mol (ppb) |
Range Noise | ≤ 5 nmol/mol (ppb) |
Detection Limit | ≤ 2nmol/mol (ppb) |
Zero Drift | ±5 nmol/mol (ppb) / 24 hours |
20% Range Drift | ±5 nmol/mol (ppb) / 24 hours |
80% Range Drift | ±10 nmol/mol (ppb) / 24 hours |
Long-term Zero Drift | ±10 nmol/mol (ppb) / 7 days |
Long-term Range Drift | ±20 nmol/mol (ppb) / 7 days |
Indication Error | < 2 % FS |
20% Range Precision | ≤ 5 nmol/mol (ppb) |
80% Range Precision | ≤ 10 nmol/mol (ppb) |
Response Time | ≤ 300 s |
Sample Flow Rate | 800 sccm ±10 % |
Operating Temperature | 5 °C-40 °C |
Operating Humidity | 0-95 % RH |
Dimensions | 432 mm x 620 mm x 178mm |
Weight | < 20 kg |
Power Supply | AC 220+22 V , 50+1 Hz |