LGH-240 Ozone Analyzer

LGH-240 Ozone Analyzer

The ozone analyzer is developed for ambient air quality monitoring. Based on UV absorption technology and advanced microprocessor technology, it provides precise and reliable measurements of O₃ in the range of μmol/mol-umol/mol (ppb to ppm).

Overview

The ozone analyzer is developed for ambient air quality monitoring. Based on UV absorption technology and advanced microprocessor technology, it provides precise and reliable measurements of O3 in the range of μmol/mol-umol/mol (ppb to ppm).


This product offers high accuracy, reliability, quick response, easy operation, and broad applicability. The analyzer supports both manual and automatic calibration, with customizable settings. It also allows for remote control and data upload. Real-time measurements and instrument status can be automatically transmitted and accessed, enabling user to monitor O3 instantly and accurately.

Features

· UV absorption method measurement, meeting EPA standards.

· Measurement range: 0-500 ppb and 0-20ppm, adjustable by the user.

· Microprocessor-driven for multi-functional operation.

· Continuous self-check with alarm function.

· Light source intensity self-check.

· Bi-directional RS232 port for remote control.

· Digital output for operating parameters.

· Adaptive signal filtering for faster response.

· Temperature and pressure compensation.

· Color touchscreen display.

· Large memory for automatic historical data storage.

Technical Specifications



Parameters

Specification

Range

Min: 0-500 nmol/mol (ppb)  Max0-20 umol/mol (ppm)

Units

nmol/mol (ppb),mol/mol (ppm),g/m3, mg/m2 ( Optional)

Zero Noise

≤ 1 nmol/mol (ppb)

Range Noise

≤ 5 nmol/mol (ppb)

Detection Limit

≤ 2nmol/mol (ppb)

Zero Drift

±5 nmol/mol (ppb) / 24 hours

20% Range Drift

±5 nmol/mol (ppb) / 24 hours

80% Range Drift

±10 nmol/mol (ppb) / 24 hours

Long-term Zero Drift

±10 nmol/mol (ppb) / 7 days

Long-term Range Drift

±20 nmol/mol (ppb) / 7 days

Indication Error

< 2 % FS

20% Range Precision

≤ 5 nmol/mol (ppb)

80% Range Precision

≤ 10 nmol/mol (ppb)

Response Time

≤ 300 s

Sample Flow Rate

800 sccm ±10 %

Operating Temperature

5 °C-40 °C

Operating Humidity

0-95 % RH

Dimensions

432 mm x 620 mm x 178mm

Weight

< 20 kg

Power Supply

AC 220+22 V , 50+1 Hz