LGH-230 Carbon Monoxide Analyzer

LGH-230 Carbon Monoxide Analyzer

The carbon monoxide analyzer is developed for ambient air quality monitoring . Based on gas filter correlation infrared absorption method and advanced microprocessor technology, it provides precise and reliable measurements of co in the range ofnmol/mol-umol/mol (ppb to ppm).

Overview

The carbon monoxide analyzer is developed for ambient air quality monitoring. Based on gas filter correlation infrared absorption method and advanced microprocessor technology, it provides precise and reliable measurements of co in the range ofnmol/mol-umol/mol (ppb to ppm).


This product offers high accuracy, reliability, quick response, easy operation, and broad applicability, The analyzer supports both manual and automatic calibration, with customizable settings, lt also allows for remote control and data upload. Real-time measurements and instrument status can be automatically transmitted and accessed, enabling user to monitor CO instantly and accurately.

Features

· Infrared absorption measurement,meeting EPA standards.

· Measurement range:0-50ppm and 0-1000ppm,adjustable by the user.

· Microprocessor-driven for multi-functional operation.

· Continuous self-check with alarm function.

· Light source intensity self-check.

· Bi-directional RS232 port for remote control.

· Digital output for operating parameters.

· Adaptive signal filtering for faster response.

· Temperature and pressure compensation.

· Color touchscreen display.

· Large memory for automatic historical data storage.


Technical Specifications



Parameters

Specification

Range

0-500 μmol/mol (ppb) ,0-1000 umol/mol (ppm)

Units

μmol/mol (ppb),ug/m(optional)

Zero Noise

≤ 0.25 μmol/mol (ppb)

Range Noise

≤ 1 μmol/mol (ppb)

Detection Limit

≤ 0.5 μmol/mol (ppb)

Zero Drift

±1 μmol/mol (ppb) / 24 hours

20% Range Drift

±1 μmol/mol (ppb) / 24 hours

80% Range Drift

±11 μmol/mol (ppb) / 24 hours

Long-term Zero Drift

±2 μmol/mol (ppb) / 7 days

Long-term Range Drift

±2 μmol/mol (ppb) / 7 days

Indication Error

< 2 % FS

20% Range Precision

≤ 0.5 μmol/mol (ppb)

80% Range Precision

≤ 0.5 μmol/mol (ppb)

Response Time

≤ 240 s

Sample Flow Rate

800±150 sccm

Operating Temperature

-10 °C-40 °C

Operating Humidity

0-85 % RH, no condensation

Dimensions

432 mm x 630 mm x 178 mm

Weight

< 20 kg

Power Supply

AC 220±22 V , 50±1 Hz